Section 8.  Operation 
 
 
 Offset Voltage Compensation Options 
CRBasic 
Measurement 
Instruction 
Input Reversal 
(RevDiff =True) 
Excitation 
Reversal 
(RevEx = True) 
Measure 
Offset During 
Measurement 
(MeasOff = True) 
Measure Offset 
During 
Background 
Calibration 
(RevDiff = False) 
(RevEx = False) 
(MeasOff = False) 
AM25T() 
   
 
 
BrHalf() 
 
 
 
 
BrHalf3W() 
 
 
 
 
BrHalf4W() 
   
 
 
BrFull() 
   
 
 
BrFull6W() 
   
 
 
TCDiff() 
 
 
 
 
TCSe() 
 
 
   
Therm107() 
 
 
 
 
Therm108() 
 
 
 
 
Therm109() 
 
 
 
 
VoltDiff() 
 
 
 
 
VoltSe() 
 
 
   
 
 
Input and Excitation Reversal 
Reversing inputs (differential measurements) or reversing polarity of excitation 
voltage (bridge measurements) cancels stray voltage offsets.    For example, if 3 
µV offset exists in the measurement circuitry, a 5 mV signal is measured as 5.003 
mV.  When the input or excitation is reversed, the second sub-measurement is –
4.997 mV.    Subtracting the second sub-measurement from the first and then 
dividing by 2 cancels the offset: 
5.003 mV – (–4.997 mV) = 10.000 mV 
10.000 mV / 2 = 5.000 mV 
When the CR800 reverses differential inputs or excitation polarity, it delays the 
same settling time after the reversal as it does before the first sub-measurement. 
So, there are two delays per measurement when either RevDiff or RevEx is used.   
If both RevDiff and RevEx are True, four sub-measurements are performed; 
positive and negative excitations with the inputs one way and positive and 
negative excitations with the inputs reversed.    The automatic procedure then is as 
follows, 
1.    Switches to the measurement terminals 
2.  Sets the excitation, and then settle, and then measure