Section 5.    Overview 
 
 
low signal is simply sensor ground (0 mV).    A single-ended measurement 
measures the high signal with reference to ground, with the low signal tied to 
ground.    A differential measurement measures the high signal with reference to 
the low signal.    Each configuration has a purpose, but the differential 
configuration is usually preferred. 
A differential configuration may significantly improve the voltage measurement.   
Following are conditions that often indicate that a differential measurement 
should be used: 
•  Ground currents cause voltage drop between the sensor and the signal-
ground terminal.    Currents >5 mA are usually considered undesirable.   
These currents may result from resistive-bridge sensors using voltage 
excitation, but these currents only flow when the voltage excitation is 
applied.    Return currents associated with voltage excitation cannot 
influence other single-ended measurements of small voltage unless the 
same voltage-excitation terminal is enabled during the unrelated 
measurements. 
•  Measured voltage is less than 200 mV. 
 
FIGURE 18:   Analog Sensor Wired to 
Single-Ended Channel #1