Section 8.  Operation 
 
 
 
 
 Three Specifications Differing Between P and C Terminals 
 
P Terminal  C Terminal 
High-Frequency 
Maximum 
250 kHz  400 kHz 
Input Voltage 
Maximum 
20 Vdc  16 Vdc 
State Transition 
Thresholds 
Count upon transition 
from 
<0.9 Vdc to >2.2 Vdc 
Count upon transition 
from 
<1.2 Vdc to >3.8 Vdc 
 
 
8.1.3.8.2 Input Filters and Signal Attenuation 
P and C terminals configured for pulse input have internal filters that reduce 
electronic noise, which can cause false counts. However, input filters attenuate 
(reduce) the amplitude (voltage) of the signal.    Attenuation is a function of the 
frequency of the signal. Higher-frequency signals are attenuated more. If a signal 
is attenuated enough, it may not pass the detection thresholds required by the 
pulse count circuitry. 
The metric for filter effectiveness is τ, the filter time constant.    The higher the τ 
value, the less noise that gets through the filter.    But, the higher the τ value, the 
lower the signal frequency must be to pass the detection thresholds. 
Detection thresholds, τ values, and low-level ac pulse input ranges are listed in 
TABLE: Time Constants 
(p. 380) 
A deduction from the specifications is that while a C terminal measured with the 
TimerIO() frequency measurement may be superior for clean signals, a P 
terminal filter (much higher τ) may be required to get a measurement on an 
electronically noisy signal. 
SPEC For example, increasing voltage is required for low-level ac inputs to 
overcome filter attenuation on P terminals configured for low-level ac: 8.5 ms 
time constant filter (19 Hz 3 dB frequency) for low-amplitude signals; 1 ms time 
constant (159 Hz 3 dB frequency) for larger (> 0.7 V) amplitude signals. 
For example, the amplitude reduction that results from τ in high-frequency pulse 
input mode is illustrated in figure FIGURE: Amplitude Reduction of Pulse Count 
Waveform
 (p. 381).