EasyManua.ls Logo

Campbell CR800 Series - PT100 Brhalf3 W() Three-Wire Half-Bridge Measurement

Campbell CR800 Series
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 7. Installation
269
PT100 BrHalf3W() Three-Wire Half-Bridge Measurement
'This program example demonstrates the measurement of a 100-ohm PRT (PT100) in a three-wire
'half bridge with voltage excitation. See adjacent procedure and schematic.
'Declare constants and variables:
Const Rf = 10000000 'Value of bridge resistor
Const RS0 = 100000 'Resistance of PT100 at 0 °C from calibration program
Public X 'Raw output from the bridge
Public RS 'Calculated PT100 resistance
Public RS_RS0 'Calculated ratio RS/RS0
Public DegC 'Calculated temperature
BeginProg
Scan(1,Sec,0,0)
....'Measure X:
'BrHalf3W(Dest,Reps,Range,SEChan,ExChan,MeasPEx,ExmV,RevEx,SettlingTime,
' Integ,Mult,Offset)
BrHalf3W(X,1,mV25,1,Vx1,1,2033,True,0,250,1,0)
'Calculate RS and RS_RS0:
RS = X * Rf
RS_RS0 = RS/RS0
....'Calculate temperature from RS_RS0:
'PRTCalc(Dest,Reps,Source,PRTType,Mult,Offset)
PRTCalc(DegC,1,RS_RS0,1,1.0,0)
NextScan
EndProg
Notes
The three-wire half-bridge compensates for lead-wire resistance by
assuming that the resistance of wire a is the same as the resistance of
wire b (see FIGURE: PT100 BrHalf3W() Three-Wire Half-Bridge
Schematic
(p. 266)). The maximum difference expected in wire resistance
is 2%, but is more likely to be on the order of 1%.
The average resistance of 22 AWG wire is 16.5 Ω per 1000 feet, which
would give 500 ft lead wires (for example) a nominal resistance of 8.3 Ω.
Two percent of 8.3 Ω is 0.17 Ω. Assuming that the greater resistance is
in wire b, the resistance measured for the PRT in the ice bath (RS0) is
100.17 Ω, and the resistance at 40 °C (RS) is 115.71 Ω.
At 40 °C, because of the error from wire b, the measured ratio RS/RS0 is
1.1551 while the ratio without the error would be 115.54/100 = 1.1554.
As a result, the temperature computed by PRTCalc() from the ratio with
the error is about 0.43 °C higher than the temperature measured without
the error from wire b. This source of error does not exist in a four-wire
half-bridge configuration.

Table of Contents

Related product manuals