Section 7.  Installation 
 
 
7.7.16.5 Example: 100 Ω PRT in Four-Wire Full Bridge with Voltage 
Excitation (PT100 / BrFull() ) 
FIGURE 67:   PT100 BrFull() Four-Wire Full-Bridge Schematic 
 
 
Procedure 
1. Build circuit
1
: 
a. Use FIGURE: PT100 BrFull() Four-Wire Full-Bridge Schematic 
(p. 270) as a 
template. 
b. Choose a 1%, 10 ppm/°C, 5000000 Ω (5 kΩ) resistors for R1 and R4 
c. Balance the bridge. 
i. Find the midpoint of the temperature range.    The range of –40 to 60 
°C is selected for this procedure, so the midpoint is 10 °C. 
ii. Select a 1% resistor for R2 with a resistance that is approximately 
equal to the resistance of the PRT at 10 °C.    See Procedure Information 
(PT100 BrFull() Full Bridge).    Since a 103.9 Ω resistor is hard to find, 
use a 100 Ω resistor.   It is close enough.    Use 5 ppm/°C resistors.   
Frequently, all the resistors in a full bridge are submerged in the medium 
to be measured, so they may see large temperature changes.    5 ppm 
resistors are more thermally stable than 10 ppm resistors. 
2. Wire circuit to datalogger: 
Use FIGURE: PT100 BrFull() Four-Wire Full Bridge Schematic 
(p. 270) as the 
wiring diagram. 
3. Calculate excitation voltage: 
Use the following equation to calculate the best excitation voltage (VX) for 
the measurement range –40 to 60 °C.    The equation reduces the absolute 
result by 1% to allow for resistor inaccuracy: 
VX
max
 = (VS
max
 / ((RS
max
 / (RS
max
 + R4)) – (R2 / R1 + R2))) • 0.99