EasyManua.ls Logo

Campbell CR800 Series - Troubleshooting - Miscellaneous Errors; Voltage Calibration Error

Campbell CR800 Series
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 10. Troubleshooting
487
the Compile, Save and Send feature of older versions of CRBasic
Editor.
o A new program (even the same program) was inadvertently sent to
the CR800 through the Connect client or Set Up client in LoggerNet.
o The program was stopped through datalogger support software File
Control or LoggerLink software.
The CPU: drive was inadvertently formated.
A network peripheral (NL115, NL120, NL200, or NL240) was added to
the CR800 when there was previously no network peripheral, and so
forced the CR800 to reallocate memory.
A hardware failure, such as memory corruption, occurred.
Inserting or removing memory cards will generally do nothing to cause
the CR800 to miss data. These events affect table definitions because
they can affect table size allocations, but they will not create a situation
where data recovery is necessary.
Data can usually be recovered using the Datalogger Data Recovery wizard
available in DevConfig
(p. 103). Recovery is possible because data in memory is
not usually destroyed, only lost track of. So, the wizard recovers "data" from the
entire memory, whether or not that memory has been written to, or written to
recently.
Once you have run through the recovery procedure, consider the following:
If a CRD: drive (memory card) or a USB: drive (Campbell Scientific mass storage
device) has been removed since the data was originally stored, then the
Datalogger Data Recovery is run, the memory pointer will likely be in the wrong
location, so the recovered data will be corrupted. If this is the case, put the CRD:
or USB: drive back in place and re-run the Datalogger Data Recovery wizard
before restarting the CRBasic program.
In any case, even when the recovery runs properly, the result will be that good
data is recovered mixed with sections of empty or old junk. With the entire data
dump in one file, you can sort through the good and the bad.
10.13 Troubleshooting Miscellaneous Errors
10.13.1 Voltage Calibration Error!
An input to an analog channel maybe outside ±8 Vdc:
Use a volt meter to check between each analog input terminal and a
ground terminal that analog inputs are not greater than ±5 Vdc.
Check for condensation which may cause leakage of 12 Vdc into other
regions of CR800 circuitry.

Table of Contents

Related product manuals