Section 7.  Installation 
 
 
 Scan(100,mSec,100,0) 
 'Measure Bridge Resistance 
 BrFull(Raw_mVperV,1,mV25,1,Vx1,1,2500,True ,True ,0,250,1.0,0) 
 
 'Calculate Strain for 1/4 Bridge (1 Active Element) 
 StrainCalc(microStrain,1,Raw_mVperV,Zero_mVperV,1,GF_Adj,0) 
 
 'Steps (1) & (3): Zero Calibration 
 'Balance bridge and set Zero_Mode = 1 in numeric monitor. Repeat after 
 'shunt calibration. 
 FieldCalStrain(10,Raw_mVperV,1,0,Zero_mVperV,Zero_Mode,0,1,10,0 ,microStrain) 
 'Step (2) Shunt Calibration 
 'After zero calibration, and with bridge balanced (zeroed), set  
 'KnownRes = to gage resistance (resistance of gage at rest), then set  
 'Shunt_Mode = 1. When Shunt_Mode increments to 3, position shunt resistor 
 'and set KnownRes = shunt resistance, then set Shunt_Mode = 4. 
 FieldCalStrain(13,MicroStrain,1,GF_Adj,0,Shunt_Mode,KnownRes,1,10,GF_Raw,0) 
 
 CallTable CalHist 
 NextScan 
 
 
7.7.11.6.3  FieldCalStrain() Quarter-Bridge Shunt Example 
With CRBasic example FieldCalStrain() Calibration (p. 230) sent to the CR800, and 
the strain gage stable, use the CR1000KD Keyboard/Display or software numeric 
monitor to change the value in variable KnownRes to the nominal resistance of 
the gage, 1000 Ω, as shown in figure Strain Gage Shunt Calibration Start
 (p. 231). 
Set Shunt_Mode to 1 to start the two-point shunt calibration. When Shunt_Mode 
increments to 3, the first step is complete. 
To complete the calibration, shunt R1 with the 249 kΩ resistor. Set variable 
KnownRes to 249000.    As shown in figure Strain Gage Shunt Calibration Finish
 
(p. 232),
 set Shunt_Mode to 4. When Shunt_Mode = 6, shunt calibration is 
complete. 
 
FIGURE 60:   Strain Gage Shunt 
Calibration Start