Section 7.  Installation 
 
 
 'PERFORM A MULTIPLIER CALIBRATION. 
 'Start by setting variable CalMode = 1. Finished when variable CalMode = 6. 
 'FieldCal(Function, MeasureVar, Reps, MultVar, OffsetVar, Mode, KnownVar, Index, Avg) 
 FieldCal(3,WC,1,WCMultiplier,0,CalMode,KnownWC,1,30) 
 
 'If there was a calibration, store it into data table CalHist 
 CallTable(CalHist) 
 
 NextScan 
 
 
7.7.11.5.5  FieldCal() Zero Basis (Opt 4) Example 
Zero-basis calibration (FieldCal() instruction Option 4) is designed for use in 
static vibrating wire measurements.    For more information, refer to these 
manuals available at www.campbellsci.com: 
AVW200-Series Two-Channel VSPECT Vibrating Wire Measurement Device 
CR6 Measurement and Control Datalogger Operators Manual 
 
7.7.11.6 Field Calibration Strain Examples 
Related Topics: 
 •  Strain Measurements — Overview (p. 70)   
 •  Strain Measurements — Details
 (p. 343)   
 •  FieldCalStrain() Examples
 (p. 228)   
Strain-gage systems consist of one or more strain gages, a resistive bridge in 
which the gage resides, and a measurement device such as the CR800 datalogger.   
The FieldCalStrain() instruction facilitates shunt calibration of strain-gage 
systems and is designed exclusively for strain applications wherein microstrain is 
the unit of measure.    The FieldCal() instruction (see Field Calibration Examples 
(p. 217)) is typically used in non-microstrain applications. 
Shunt calibration of strain-gage systems is common practice.    However, the 
technique provides many opportunities for misapplication and misinterpretation.   
This section is not intended to be a primer on shunt-calibration theory, but only to 
introduce use of the technique with the CR800 datalogger.    Campbell Scientific 
strongly urges users to study shunt-calibration theory from other sources.    A 
thorough treatment of strain gages and shunt-calibration theory is available from 
Vishay using search terms such as 'micro-measurements', 'stress analysis', 'strain 
gages', 'calculator list', at: 
http://www.vishaypg.com 
 
7.7.11.6.1  FieldCalStrain() Shunt Calibration Concepts 
1.  Shunt calibration does not calibrate the strain gage itself. 
2.  Shunt calibration does compensate for long leads and non-linearity in the 
resistive bridge. Long leads reduce sensitivity because of voltage drop.   
FieldCalStrain() uses the known value of the shunt resistor to adjust the gain 
(multiplier / span) to compensate.    The gain adjustment (S) is incorporated by