EasyManua.ls Logo

Campbell CR800 Series - Page 263

Campbell CR800 Series
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 7. Installation
263
b. Rf should approximately equal the resistance of the PT100 at 0 °C. Use a
1%, 10 ppm/°C resistor.
2. Wire circuit to datalogger:
Use FIGURE: PT100 BrHalf4W() Four-Wire Half-Bridge Schematic
(p. 262) as
the wiring diagram.
3. Calculate excitation voltage
2
:
Use the following equation to calculate the best excitation voltage (VX) for
the measurement range 40 to 60 °C. The equation reduces the absolute
result by 1% to allow for resistor inaccuracy:
VX
max
= (VS
max
• (Rf
+ RS
max
+ R1) / RS
max
) • 0.99
where,
VS
max
= 25 mV (maximum voltage in the ±25 mV input range)
Rf = 100000 mΩ (100 Ω)
R1 = 10000000 mΩ (10 kΩ)
RS
max
= 123240 mΩ (PT100 at 60 °C)
3
so,
VX
max
= 2053 mV
4. Calibrate the PT100:
If the PRT accuracy specification is good enough, and you trust it, assume
RS
0
= 100000 mΩ. Otherwise, do the following procedure:
a. Enter CRBasic EXAMPLE: PT100 BrHalf4W() Four-Wire Half-Bridge
Calibration
(p. 264) into the CR800. It is already programmed with the
excitation voltage from step 3.
b. Place the PRT in an ice bath (0 °C).
c. Measure the PRT. If you are doing a dry run, assume the result of
BrHalf4W() = X
0
= 0.01000.
d. Calculate RS
0
RS
0
= X
0
• Rf = 100000 mΩ
Wow! We are lucky to have a perfect PRT! In the real world, PRT
resistance at 0 °C will probably land on either side of 100 Ω.
5. Measure the sensor:

Table of Contents

Related product manuals