Section 8.  Operation 
 
 
performed as part of the routine auto-calibration of the CR800.  Single-ended 
measurement instructions VoltSE() and TCSe() MeasOff parameter determines 
whether the offset voltage measured is done at the beginning of measurement 
instruction, or as part of self-calibration. This option provides you with the 
opportunity to weigh measurement speed against measurement accuracy.    When 
MeasOff = True, a measurement of the single-ended offset voltage is made at the 
beginning of the VoltSE() instruction. When MeasOff = False, an offset voltage 
measurement is made during self-calibration.    For slowly fluctuating offset 
voltages, choosing MeasOff = True for the VoltSE() instruction results in better 
offset voltage performance. 
Ratiometric measurements use an excitation voltage or current to excite the sensor 
during the measurement process.    Reversing excitation polarity also reduces 
offset voltage error.    Setting the RevEx parameter to True programs the 
measurement for excitation reversal.    Excitation reversal results in a polarity 
change of the measured voltage so that two measurements with opposite polarity 
can be subtracted and divided by 2 for offset reduction similar to input reversal for 
differential measurements.    Ratiometric differential measurement instructions 
allow both RevDiff and RevEx to be set True.  This results in four measurement 
sequences:   
•  positive excitation polarity with positive differential input polarity 
•  negative excitation polarity with positive differential input polarity 
•  positive excitation polarity with negative differential input polarity 
•  positive excitation polarity then negative excitation differential input 
polarity 
For ratiometric single-ended measurements, such as a BrHalf(), setting RevEx = 
True results in two measurements of opposite excitation polarity that are 
subtracted and divided by 2 for offset voltage reduction.    For RevEx = False for 
ratiometric single-ended measurements, an offset-voltage measurement is made 
during the self-calibration. 
When analog voltage signals are measured in series by a single measurement 
instruction, such as occurs when VoltSE() is programmed with Reps = 2 or more, 
measurements on subsequent terminals may be affected by an offset, the 
magnitude of which is a function of the voltage from the previous measurement. 
While this offset is usually small and negligible when measuring large signals, 
significant error, or NAN, can occur when measuring very small signals. This 
effect is caused by dielectric absorption of the integrator capacitor and cannot be 
overcome by circuit design. Remedies include the following: 
•  Program longer settling times 
•  Use an individual instruction for each input terminal, the effect of which 
is to reset the integrator circuit prior to filtering. 
•  Avoid preceding a very small voltage input with a very large voltage 
input in a measurement sequence if a single measurement instruction 
must be used.