EasyManua.ls Logo

Campbell CR800 Series - Page 347

Campbell CR800 Series
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 8. Operation
347
Note This section contains advanced information not required for normal
operation of the CR800.
Summary
Voltage input limits for measurement are ±5 Vdc. Input Limits is the
specification listed in Specifications
(p. 91).
• Common-mode range is not a fixed number. It varies with respect to
the magnitude of the input voltage.
The CR800 has features that help mitigate some of the effects of
signals that exceed the Input Limits specification or the common-mode
range.
With reference to the figure PGIA with Input-Signal Decomposition (p. 348), the
PGIA processes the voltage difference between V+ and V. It ignores the
common-mode voltage, or voltages that are common to both inputs. The figure
shows the applied input voltage decomposed into a common-mode voltage (V
cm
)
and the differential-mode component (V
dm
) of a voltage signal. V
cm
is the
average of the voltages on the V+ and Vinputs. So, V
cm
= (V+ + V)/2 or the
voltage remaining on the inputs when V
dm
= 0. The total voltage on the V+ and
V– inputs is given as V+ = V
cm
+ V
dm
/2, and V= V
cm
– V
dm
/2, respectively.
The PGIA ignores or rejects common-mode voltages as long as voltages at V+
and Vare within the Input Limits specification, which for the CR6 is ±5 Vdc
relative to ground. Input voltages wherein V+ or V, or both, are beyond the ±5
Vdc limit may suffer from undetected measurement errors. The Common-Mode
Range defines the range of common-mode voltages that are not expected to
induce undetected measurement errors. Common-Mode Range is different than
Input Limits when the differential mode voltage in non-negligible. The following
relationship is derived from the PGIA figure as:
Common-Mode Range = ±5 Vdc |V
dm
/2|.
The conclusion follows that the common-mode range is not a fixed number, but
instead decreases with increasing differential voltage. For differential voltages
that are small compared to the input limits, common-mode range is essentially
equivalent to Input Limits. Yet for a 5000 mV differential signal, the common-
mode range is reduced to ±2.5 Vdc, whereas Input Limits are always ±5 Vdc.
Consequently, the term Input Limits is used to specify the valid voltage range of
the V+ and Vinputs into the PGIA.

Table of Contents

Related product manuals