Section 8.  Operation 
 
 
 •  Excitation reversal (RevEx = True) 
 •    Longer settling times 
Voltage offset can be the source of significant error.    For example, an offset of 3 
μV on a 2500 mV signal causes an error of only 0.00012%, but the same offset on 
a 0.25 mV signal causes an error of 1.2%.    The primary sources of offset voltage 
are ground currents and the Seebeck effect. 
Single-ended measurements are susceptible to voltage drop at the ground terminal 
caused by return currents from another device that is powered from the CR800 
wiring panel, such as another manufacturer's comms modem, or a sensor that 
requires a lot of power.    Currents >5 mA are usually undesirable.    The error can 
be avoided by routing power grounds from these other devices to a power ground 
G terminal on the CR800 wiring panel, rather than using a signal ground ( ) 
terminal.    Ground currents can be caused by the excitation of resistive-bridge 
sensors, but these do not usually cause offset error.    These currents typically only 
flow when a voltage excitation is applied.    Return currents associated with 
voltage excitation cannot influence other single-ended measurements because the 
excitation is usually turned off before the CR800 moves to the next measurement.   
However, if the CRBasic program is written in such a way that an excitation 
terminal is enabled during an unrelated measurement of a small voltage, an offset 
error may occur. 
The Seebeck effect results in small thermally induced voltages across junctions of 
dissimilar metals as are common in electronic devices.    Differential 
measurements are more immune to these than are single-ended measurements 
because of passive voltage cancelation occurring between matched high and low 
pairs such as 1H/1L.    So use differential measurements when measuring critical 
low-level voltages, especially those below 200 mV, such as are output from 
pyranometers and thermocouples.    Differential measurements also have the 
advantage of an input reversal option, RevDiff.  When RevDiff is True, two 
differential measurements are made, the first with a positive polarity and the 
second reversed.    Subtraction of opposite polarity measurements cancels some 
offset voltages associated with the measurement. 
Single-ended and differential measurements without input reversal use an offset 
voltage measurement with the PGIA inputs grounded.  For differential 
measurements without input reversal, this offset voltage measurement is 
performed as part of the routine auto-calibration of the CR800.    Single-ended 
measurement instructions VoltSE() and TCSe() MeasOff parameter determines 
whether the offset voltage measured is done at the beginning of measurement 
instruction, or as part of self-calibration. This option provides you with the 
opportunity to weigh measurement speed against measurement accuracy.    When 
MeasOff = True, a measurement of the single-ended offset voltage is made at the 
beginning of the VoltSE() instruction. When MeasOff = False, an offset voltage 
measurement is made during self-calibration.    For slowly fluctuating offset 
voltages, choosing MeasOff = True for the VoltSE() instruction results in better 
offset voltage performance. 
Ratiometric measurements use an excitation voltage or current to excite the sensor 
during the measurement process.    Reversing excitation polarity also reduces 
offset voltage error.    Setting the RevEx parameter to True programs the 
measurement for excitation reversal.    Excitation reversal results in a polarity 
change of the measured voltage so that two measurements with opposite polarity