5.3 Configuring Basic Settings for Measurement
86
Maximum Recording Period for Each Recording Interval
Measure continuously without specifying a recording period.
1. Set [Continuous] on the measurement settings page to ON.
Each click of the button toggles the setting ON and OFF.
In the case of normal sampling and dual sampling
2. Start measurement to measure continuously until measurement is
stopped.
(Stop trigger, when repeat recording is ON)
Recording
Interval
(High-speed)
Maximum Recording
Period
Recording
Interval
(High-speed)
Maximum Recording
Period
Recording
Interval
(High-speed)
Maximum Recording
Period
10 ms
1 day
1 s 100 days 1 m 500 days
20 ms 2 days 2 s 200 days 2 m 500 days
50 ms 5 days 5 s 500 days 5 m 500 days
100 ms 10 days 10 s 500 days 10 m 500 days
200 ms 20 days 20 s 500 days 20 m 500 days
500 ms 50 days 30 s 500 days 30 m 500 days
1 h 500 days
Continuous Recording
Time
Recording operation
Start button
Recording operation
Recording operation
Start measurement
Data timer for saving
T T
Time
Trigger Trigger
Start button