Appendix
D: Convert Model 2400 to Model 2461 code Model 2461 Interactive SourceMeter® Instrument
Reference Manual
D-30 2461-901-01 A/November 2015
Model 2400 command
:ARM[:SEQuence[1]][:LAYer[1]][:TCONfigure]:OLINe
:ARM[:SEQuence[1]][:LAYer[1]][:TCONfigure]:OLINe?
Model 2461 command
Not available
Notes
For similar functionality, use a digital I/O line with the trigger model.
Generate a notify event in the trigger model that feeds the stimulus setting to a
digital I/O line to pulse as needed in the trigger model.
See:
:TRIGger:BLOCk:NOTify (on page 6-206)
:TRIGger:DIGital<n>:OUT:STIMulus (on page 6-214)
Model 2400 command
:ARM[:SEQuence[1]][:LAYer[1]][:TCONfigure]:OUTPut
:ARM[:SEQuence[1]][:LAYer[1]][:TCONfigure]:OUTPut?
Model 2461 command
Not available
Notes
Similar functionality available with :TRIGger:BLOCk:NOTify (on page 6-206) set
to a digital I/O line.
Model 2400 command
:ARM[:SEQuence[1]][:LAYer[1]]:TIMer
:ARM[:SEQuence[1]][:LAYer[1]]:TIMer?
Model 2461 command
Not available
Notes
For similar functionality, use :TRIGger:TIMer<n>:COUNt (on page 6-236) and
:TRIGger:BLOCk:WAIT (on page 6-209).
Model 2400 command
:INITiate[:IMMediate]
Model 2461 command
:INITiate[:IMMediate] (on page 6-182)
Notes
Model 2400 command
:TRIGger:CLEar
Model 2461 command
Not available
Notes
Model 2400 command
:TRIGger[:SEQuence[1]]:COUNt
:TRIGger[:SEQuence[1]]:COUNt?
Model 2461 command
Not available
Notes
For similar functionality, use :TRIGger:BLOCk:BRANch:COUNter (on page 6-
186). You can also use the Simple Loop trigger model; for information, refer to
Predefined trigger models (on page 3-125).
Model 2400 command
:TRIGger[:SEQuence[1]]:DELay
:TRIGger[:SEQuence[1]]:DELay?
Model 2461 command
Not available
Notes
For similar functionality, use :TRIGger:BLOCk:DELay:CONStant (on page 6-
199).
Model 2400 command
:TRIGger[:SEQuence[1]]:SOURce
:TRIGger[:SEQuence[1]]:SOURce?
Model 2461 command
Not available
Notes
Model 2400 command
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:INPut
:TRIGger[:SEQuence[1]][:TCONfigure][:ASYNchronous]:INPut?
Model 2461 command
Not available
Notes

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals