Interactive SourceMeter® Instrument Reference Manual Appendix D: Convert Model 2400 to Model 2461 cod
2461-901-01 A/November 2015 D-31
Model 2400 command
:TRIGger[:SEQuence[1]][:TCONfigure]:DIRection
:TRIGger[:SEQuence[1]][:TCONfigure]:DIRection?
Model 2461 command
Notes
Similar functionality available with :TRIGger:BLOCk:WAIT (on page 6-209).
Model 2400 command
:TRIGger[:SEQuence[1]][:TCONfigure]:ILINe
:TRIGger[:SEQuence[1]][:TCONfigure]:ILINe?
Model 2461 command
Notes
For similar functionality, use a digital I/O line with the trigger model.
Generate a notify event in the trigger model that feeds the stimulus setting to a
digital I/O line to pulse as needed in the trigger model.
See:
• :TRIGger:BLOCk:NOTify (on page 6-206)
•
:TRIGger:DIGital<n>:OUT:STIMulus (on page 6-214)
Model 2400 command
:TRIGger[:SEQuence[1]][:TCONfigure]:OLINe
:TRIGger[:SEQuence[1]][:TCONfigure]:OLINe?
Model 2461 command
Notes
For similar functionality, use a digital I/O line with the trigger model.
Generate a notify event in the trigger model that feeds the stimulus setting to a
digital I/O line to pulse as needed in the trigger model.
See:
• :TRIGger:BLOCk:NOTify (on page 6-206)
•
:TRIGger:DIGital<n>:OUT:STIMulus (on page 6-214)
Model 2400 command
:TRIGger[:SEQuence[1]][:TCONfigure]:OUTPut
:TRIGger[:SEQuence[1]][:TCONfigure]:OUTPut?
Model 2461 command
Notes
Similar functionality available with :TRIGger:BLOCk:NOTify (on page 6-206) set
to a digital I/O line.
Model 2400 command
:TRIGger:SEQuence2:SOURce
:TRIGger:SEQuence2:SOURce?
Model 2461 command
Notes
Model 2400 command
:TRIGger:SEQuence2:TOUT
:TRIGger:SEQuence2:TOUT?
Model 2461 command
Notes