Using SCPI commands to select a reading buffer:
To make a measurement and store the readings in a specific reading buffer, send the command:
:READ? "<bufferName>"
If you do not specify a buffer name, readings are stored in defbuffer1.
An alternative to sending the :READ? "<bufferName>" command is to send the command:
:TRACe:TRIGger "<bufferName>"
The :TRACe:TRIGger command stores readings in the specified reading buffer. If no buffer is
specified for the parameter, defbuffer1 is used. To see the readings stored in the buffer after using
this command, use the :FETCh? command to see the last reading stored in the buffer or the
:TRACe:DATA? command to see multiple readings from the buffer.
To specify a user-defined reading buffer, you must create the buffer first.
To select current as the measurement function, measure current, and return the readings in the
testData reading buffer, send the following commands:
:SENSe:FUNCtion "CURRent"
:READ? "testData"
To measure current and store the readings in the defbuffer2 reading buffer, send the following
command:
:MEASure:CURRent? "defbuffer2"
To measure voltage and store the readings in the defbuffer2 reading buffer, send the following
command:
:MEASure:VOLTage? "defbuffer2"
To measure current and return the relative time and a reading, send the following command:
:MEASure:CURRent? "testData", REL, READ
Buffer storage is consistent whenever readings are taken. Parameters such as REL and READ only
affect what is included in the response. If you do not include parameters, the command only returns
the reading.
Using TSP commands to select a reading buffer:
To make a measurement and store the readings in a specific reading buffer, use the
smu.measure.read(bufferName) function. If you do not specify a buffer when you use the
smu.measure.read() function, readings are stored in defbuffer1.
To measure voltage and store the readings in the voltMeasBuffer, send the following commands:
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.read(voltMeasBuffer)
To measure voltage, store the readings in the voltMeasBuffer, and print the last reading in the
buffer, send the following command:
print(smu.measure.read(voltMeasBuffer))

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV