Section
4: Source-measure considerations Model 2461 Interactive SourceMeter® Instrument
Reference Manual
4-20 2461-901-01 A/November 2015
Voltage measure accuracy:
Input signal = (20 x 100 mA) = 2 V
Accuracy specification of 2 V range = ± (0.012 % of output + 300 µV)
= ± {(2 V x 0.00012) + 300 µV}
= ± {240 µV + 300 µV}
= ± 540 µV
Error %
=± 0.027 %
Total measurement uncertainty = 0.035 % + 0.027 % = 0.062 %
For higher accuracy measurements when using SMU Instruments, use the source readback function
to actually measure the source output. Use the measured source value to calculate the resistance.
To calculate the total accuracy from this example using source readback, add the current and voltage
measurement accuracy specifications.
Current measure accuracy:
Input signal = 100 mA on 100 mA range

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals