Gocator Line Profile Sensors: User Manual
Gocator Web Interface • 346
Measurements, Features, and Settingss
Measurement Illustration
X
Determines the X position of the selected feature type.
Y
Determines the Y position of the selected feature type.
Z
Determines the Z position of the selected feature type.
Measurements
Type Description
Center Point
The returned position.
Features
For more information on geometric features, see Geometric Features on page 206.
Parameter Description
Source The sensor, or combination of sensors, that provides data
for the tool's measurements. For more information, see
Source on page 201.
Feature The feature the tool uses for its measurements. One of the
following:
l Average
l Median
l Centroid
l Max X
l Min X
l Max Y
l Min Y
l Max Z
l Min Z
To set the region of a feature, adjust it graphically in the
data viewer, or expand the feature using the expand
button ( ) and enter the values in the fields. For more
information on regions, see Regions on page 201.
Filters The filters that are applied to measurement values before
they are output. For more information, see Filters on page
209.
Decision The Max and Min settings define the range that determines
whether the measurement tool sends a pass or fail decision
to the output. For more information, see Decisions on page
208.
Parameters

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LMI Technologies Gocator 2100 Series and is the answer not in the manual?

LMI Technologies Gocator 2100 Series Specifications

General IconGeneral
BrandLMI Technologies
ModelGocator 2100 Series
CategoryAccessories
LanguageEnglish

Related product manuals