System Configuration
R&S
®
ESR
526User Manual 1175.7068.02 ─ 12
"Install Option by XML" on page 526
Only user accounts with administrator rights are able to install options.
Install Option ← Option Licenses
Opens an edit dialog box to enter the license key for the option that you want to install.
If an option is about to expire, a message box is displayed to inform you. You can then
use this softkey to enter a new license key.
If an option has already expired, a message box appears for you to confirm. In this
case, all instrument functions are unavailable (including remote control) until you
reboot the R&S ESR. Then use the "Install Option" softkey to enter the new license
key.
For more information about the option in question, refer to the System Info softkey in
the "Setup" menu.
Only user accounts with administrator rights are able to install options.
Install Option by XML ← Option Licenses
Opens an edit dialog to install an additional option to the R&S ESR using an XML file.
Enter or browse for the name of an XML file on the instrument that contains the option
key and press "Select".
Only user accounts with administrator rights are able to install options.
Application Setup Recovery
Controls instrument behavior when switching between measurement applications, e.g.
from "Spectrum" to "Analog Demod" and back.
If this softkey is activated, the current instrument settings are stored when you switch
to a different application. When you switch back to the previous application, the corre-
sponding instrument settings are restored. Thus, the settings of the individual applica-
tions are independent of each other.
If the softkey is deactivated (default), only a few parameters of the current instrument
setting are passed between applications (e.g. center frequency, level settings).
Note that this setting is not deactivated during a preset operation, i.e. you must deacti-
vate it manually, if necessary.
Remote command:
SYSTem:APPLication:SRECovery[:STATe] on page 920
Preset Receiver
Selects the default settings defined for the Receiver mode to be restored when you
perform an instrument preset.
Remote command:
SYSTem:PRESet:COMPatible on page 922
Preset Spectrum
Selects the default settings defined for the Spectrum mode to be restored when you
perform an instrument preset.
Remote command:
SYSTem:PRESet:COMPatible on page 922
Instrument Setup and Interface Configuration – SETUP Key

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals