Remote Control
R&S
®
ESR
689User Manual 1175.7068.02 ─ 12
Example:
INIT:CONT OFF
Switches to single sweep mode
CALC:MARK:FUNC:ZOOM 1kHz;*WAI
Activates zooming and waits for its end.
Manual operation: See "Marker Zoom (span > 0)" on page 408
CALCulate<n>:MARKer<m>:SCOupled[:STATe] <State>
This command couples or decouples the marker frequency to the scan range settings.
Parameters:
<State> ON | OFF
*RST: ON
Example:
CALC:MARK:SCO ON
Couples the scan range settings to the marker frequency.
Manual operation: See "Settings Coupled" on page 204
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT <Limit>
This command sets the left limit of the marker search range.
If the power measurement in zero span is active, this command limits the evaluation
range to the trace.
Note: The function is only available if the search limit for marker and delta marker is
switched on (see CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]).
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Parameters:
<Limit> The value range depends on the span or sweep time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
Range: 0 to MAX
*RST: left diagram border
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
Manual operation: See "Left Limit" on page 206
See "Left Limit" on page 297
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHT <Limit>
This command sets the right limit of the marker search range.
Remote Commands in Receiver Mode