Remote Control
R&S
®
ESR
835User Manual 1175.7068.02 ─ 12
[SENSe:]SWEep:TIME <Time>
This command defines the sweep time.
The range depends on the frequency span.
If this command is used in "Spectrum" mode, automatic coupling to resolution band-
width and video bandwidth is switched off.
Parameters:
<Time> refer to data sheet
*RST: (automatic)
Example:
SWE:TIME 1s
Manual operation: See "Sweep Time" on page 243
See "Sweeptime Manual" on page 362
See "Meas Time" on page 460
[SENSe:]SWEep:TIME:AUTO <State>
In realtime mode, this command automatically sets the sweep time to 32 ms.
In analyzer mode, this command controls the automatic coupling of the sweep time to
the frequency span and bandwidth settings. If [SENSe:]SWEep:TIME is used, auto-
matic coupling is switched off.
Parameters:
<State> ON | OFF
*RST: ON
Example:
SWE:TIME:AUTO ON
Activates automatic sweep time.
Manual operation: See "Sweeptime Manual" on page 362
See "Sweeptime Auto" on page 363
See "Default Coupling" on page 365
[SENSe:]SWEep:TYPE <Type>
This command selects the sweep type.
Parameters:
<Type> SWE
Selects analog frequency sweeps.
AUTO
Automatically selects the sweep type (FFT or analog frequency
sweep).
FFT
Selects FFT sweeps.
*RST: AUTO
Remote Commands in Spectrum Analyzer Mode