2-16 Rapidlab 1200 Operator’s Guide: Operating the System
02087462 Rev. V
10. Watch the sample move through the sample path and inspect the sample path for
bubbles or gaps.
Ensure that the trailing edge of the sample remains in contact with the sample/
ground temperature sensor.
Ensure that the leading edge of the sample fills the reference sensor.
Figure 2-5 Measurement Module
11. If prompted, enter demographic information and select Continue.
You can scan the patient ID and accession number in the appropriate field. For more
information about entering the data, refer to Entering Patient Sample Data‚ page 2-22.
12. View the results.
For information about how results are displayed on the screen and in printed reports,
refer to Viewing Patient Results‚ page 2-27.
13. When you finish viewing the results, select
Continue.
1 Sample/ground temperature sensor
2 Reference sensor
If... Then...
bubbles or gaps are present, select
Cancel.
no bubbles or gaps are present, select
Analyze.