Rapidlab 1200 Operator’s Guide: Troubleshooting 6-51
02087462 Rev. V
D3 Codes
D3 Slope Error: pO
2
pCO
2
D3 Slope Error: pH Na
+
K
+
Ca
++
Cl¯ Ref
Problem Solution
Sensor slope is
beyond predefined
limits during
calibration
operations.
Correct any reagent problems such as D23 codes before you
attempt to correct this D code.
Perform a 2-point calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
Check that the sensors are installed in the correct location
and are aligned, the O-rings are in place, and the
spring-loaded latch is closed.
If necessary, remove and install the sensors correctly.
If the problem persists, replace the sensor.
If the problem persists, call for technical assistance.
Problem Corrective Action
Sensor slope is
beyond predefined
limits during
calibration
operations.
Correct any reagent problems such as D23 codes before you
attempt to correct this D code.
Perform a 2-point calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
Deproteinize the sample path and condition the sensors.
Refer to Deproteinizing the Sample Path‚ page 5-4.
Check the solution level in the affected sensor and look for
bubbles.
The K
+
, Ca
++
, and Cl¯ sensors should be nearly full. The pH
and Na
+
sensors should be full. If the solution level is low,
refill the sensor.
If more than one sensor or the reference sensor has the slope
error, look for salt in the vent hole and bubbles in the sensor.
Clean and remove bubbles from the reference sensor as
needed.
Check that the sensors are installed in the correct location
and are aligned, the O-rings are in place, and the
spring-loaded latch is closed.
If necessary, remove and install the sensors correctly.
Check the sensor contacts and measurement block for salt
buildup.
Check behind the sensor for fluid leaks that can come from
the reference sensor.
If salt or leaks are present, clean the sensor with water, dry,
reinstall, and perform a 2-point calibration. Refer to
Performing Manual Calibrations‚ page 3-4.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Rapidlab 1200 and is the answer not in the manual?

Siemens Rapidlab 1200 Specifications

General IconGeneral
BrandSiemens
ModelRapidlab 1200
CategoryMedical Equipment
LanguageEnglish

Related product manuals