5-42 Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
Replacing System Components
This section contains procedures for replacing system components.
Replacing the Printer Paper
Replace the printer paper when a red stripe is displayed on the edge of the paper.
You need printer paper for this task.
1. Select
Status > Maintenance > Replacing the Printer Paper.
2. To start the procedure, select
Continue.
3. Follow the instructions in the video.
Replacing the Sample Port
Use this procedure to replace the sample port. Replace the sample port if you detect a
problem, or if the system prompts you as a result of obstructions such as fibrin clots.
You need a sample port for this task.
BIOHAZARD: Wear personal protective equipment. Use universal precautions.
Refer to Appendix A, Protecting Yourself from Biohazards for
recommended
precautions when working with biohazardous materials.
1. Select
Status > Maintenance > Replace Sample Port.
2. To start the procedure, select
Continue.
3. Follow the instructions in the video.
CAUTION: Do not pull the torn paper back through the printer. This can damage
the printing mechanism.