5-34 Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
Performing Reference Sensor Maintenance
The Reference sensor requires regular cleaning to maintain performance.
Cleaning and Inspecting the Reference Sensor
You need the following materials for this task:
•Water
Lint-free swab
1. Remove the Reference sensor.
Refer to Removing the Sensors‚ page 5-30.
Figure 5-16 Measurement Module
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
1 Reference sensor
2 Spring-loaded latch