Rapidlab 1200 Operator’s Guide: Quality Control 4-5
02087462 Rev. V
2. Introduce the QC sample:
a. Scan the barcode on the QC ampule, if necessary.
b. Invert the QC ampule several times and then open the ampule.
c. Prepare the sample.
d. Insert the sample device into the sample port.
e. Select
Analyze.
f. When prompted, remove the sample device from the sample port and then
select
Continue.
3. If prompted, enter your operator ID and then select
Continue.
4. View the results on the screen and on the printed reports.
If the Auto Print option is turned on in Setup, the system automatically prints
the QC sample report.
5. When you finish viewing the results, select
Continue.
Accessing QC Results
When the system completes QC analysis, the system displays the Results screen.
For a Required QC analysis and for Unscheduled QC analysis when you scan the
barcode on the ampule, the Results screen immediately under the banner contains
the following information:
•QC type
• Level
•Lot
For AutomaticQC samples, the system displays the level.
If the sample device is... Then...
an ampule attach a Quick adapter to the ampule.
a syringe draw the QC sample into the syringe.