5-36 Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
Filling the Reference Sensor Cassette
If you do not have a new reference sensor cassette, refill the reference sensor instead.
You need the following materials for this task:
Reference electrode refill kit
Reference electrode internal replacement kit
•Hex tool
Lint-free tissue
1. Remove the sensor.
Refer to Removing the Sensors‚ page 5-30.
2. Remove the internal reference electrode:
a. Remove the internal reference electrode using the hex tool.
Figure 5-18 Reference Sensor Cassette
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
1 Internal electrode
2 Cassette
3 Internal electrode compartment