6-38 Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
Use the following steps to perform the leak test.
1. Select Status > Diagnostics.
2. Enter your password, if necessary.
3. Select
Fluidic Functions > Leak Test.
4. Wait while the system performs the test.
a. When the screen displays a Rapidlab illustration, use it to determine the probable
area of the leak.
b. If the test fails, do the following:
look for highlighted Fluidic Function buttons
These are tests you can perform to clear the leak.
check the tubing for leaks
ensure that you installed the tubing correctly in the pinch valve
ensure that the O-rings are in place
ensure that you closed the spring-loaded latch
For Rapidlab 1245 and 1265 systems, verify that the CO-ox sample chamber
is installed correctly.
5. Select
Return.
6. To return to the Status screen, select
Return.
The system performs a wash.
Using the Pumps and Valves Screen
The Diagnostics > Pumps and Valves screen displays buttons for performing the following
actions:
Perform the Sample Pump and Wash Pump Tests
Perform the R Cartridge Valve Test
Perform the AQC Valve Test
Perform the Measurement Module Pinch Valve Test
Perform the CO-ox Pump Test