5-54 Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
Flushing the Tubing for the Samples
1. Connect the syringe tubing to the connector for samples.
2. Flush the sample tubing with a minimum of 3 mL water.
3. Flush the sample tubing with a minimum of 12 mL air.
4. Collect the water in an absorbent swab or equivalent.
Figure 5-28 Sample Tubing
1 Water collection area
2 Sample connector