Rapidlab 1200 Operator’s Guide: Troubleshooting 6-7
02087462 Rev. V
Troubleshooting Unavailable Buttons
Use this section when buttons at the Analysis screen are not available.
The Rapidlink or Rapidcomm
data management system or the
LIS connected to the
Rapidlab 1200 system turned the
parameter off.
Contact your system supervisor for assistance if
you need to analyze samples and obtain results for
this parameter.
The system could not report a
result because a parameter used
to determine the result was not
reported, not available, or sample
demographics were not entered.
Verify that the system reported valid results for all
necessary parameters.
To identify parameters that depend on other
parameters for results, refer to Required
Parameters and Sample Demographics‚ page 8-17.
If demographics editing is turned on in Setup, recall
the sample results and enter the values for the
sample demographics. Refer to Editing
Demographics‚ page 2-25.
Ensure that the parameters and sample
demographics are turned on in Setup. Refer to
Setting up Parameters‚ page 8-13.
Problem Probable Cause Corrective Action
The QC!
button on the
Analysis
screen is not
available.
The lot is expired for the
control that is currently
scheduled, or no lot is
entered for the control.
If the lot is expired, define a new lot of
control.
If no lot is defined for the current control,
enter the lot.
All parameters are out
of calibration because
they were turned off in
Setup and then turned
on again.
Perform a 2-point calibration to make the
parameter available again.
The Analyze
button has a
line through it.
If all parameters are
available, the Rapidlink
or Rapidcomm data
management system or
the LIS connected to the
Rapidlab 1200 system
has turned analysis
functions off for the
Rapidlab 1200 system.
Contact your system supervisor for
assistance if you need to analyze samples.
Table 6-5 Missing Results
Probable Cause Corrective Action