Rapidlab 1200 Operator’s Guide: Troubleshooting 6-39
02087462 Rev. V
Performing the Sample Pump and Wash Pump Tests
Use this procedure to test the sample pump or the wash pump located on the
system behind the reagent cartridge.
The pump test moves the pump at 4 speeds numbered 1 through 4. When you
sele
ct
Change, the system displays the number of the current speed on the screen.
The test is complete when the system displays speed 4 on the screen.
BIOHAZARD: Wear personal protective equipment. Use universal precautions.
Refer to Appendix A, Protecting Yourself from Biohazards for
recommended
precautions when working with biohazardous materials.
1. Select
Status > Diagnostics.
2. Enter your password, if necessary.
3. Remove the reagent cartridge:
a. Select
Cartridges > Eject R Cartridge.
b. Remove the reagent cartridge from the system and close the door.
c. Select
Return.
4. Select
Pumps and Valves.
CAUTION: Do not remove the wash or reagent cartridges for longer than 6 hours.
If the wash or reagent cartridge is removed from the system for longer than 6
hours, the cartridges are no longer valid.
To test the... Select...
sample pump, which is located on the
right as you face the system,
Sample Pump > Start.
The pump rotates toward you at the first
speed.
wash pump, which is located on the left as
you face the system,
Wash Pump > Start.
The pump rotates away from you at the
first speed.