Rapidlab 1200 Operator’s Guide: Maintenance 5-15
02087462 Rev. V
Replacing the CO-ox Waste Tubing
1. Disconnect the CO-ox waste tubing from the pump tubing and the sample
chamber connector.
Figure 5-5 CO-ox Module Tubing
2. Connect the CO-ox waste tubing to the CO-ox pump tubing and the sample
chamber connector.
1 CO-ox pump tubing
2 CO-ox waste tubing
3 Sample chamber