Rapidlab 1200 Operator’s Guide: Maintenance 5-29
02087462 Rev. V
Maintaining the Sensors
Performing regular maintenance on the sensors is important to the life of the
sensors. This section contains the following information:
Replacing the sensors
Performing reference sensor maintenance
Refilling the sensors
Replacing the Sensors
You need the following materials for this task:
Appropriate sensor
Appropriate sensor fill solution
Reference sensor replacement kit
•Hex tool
Lint-free tissue
Preparing the Sensors
Equilibrate new Glucose and Lactate biosensors at room temperature
(18° to 25°C) for at least 1 hour before installing.
Keep the biosensors in their foil packages while they are equilibrating.
BIOHAZARD: Wear personal protective equipment. Use universal precautions.
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.