EasyManua.ls Logo

Siemens Rapidlab 1200 - Using the Calibration Report to Identify Problems

Siemens Rapidlab 1200
448 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rapidlab 1200 Operator’s Guide: Troubleshooting 6-5
02087462 Rev. V
Check the solution level in the affected sensor and remove bubbles.
The K
+
, Ca
++
, and Cl¯ sensors should be nearly full. The pH and Na
+
sensors
should be completely full without any bubbles.
Check the sensor contacts and measurement block for salt buildup.
Check behind the sensors for fluid leaks that can come from the reference
sensor.
Check that the sensors are installed in the correct order and are aligned, the
O-rings are in place, and the spring-loaded latch is closed.
If the pH or Na
+
sensor have a D2 or D3, condition the sensors.
Check for cracks or leaks in the sample tubing, and replace the tubing if
necessary.
Replace the affected sensor, if necessary.
Using the Calibration Report to Identify Problems
The full calibration report shows the measurement and drift values. Rapidlab 1200
systems use a 2-point calibration curve. The two points on the curve are the Cal
Point and Slope Point.
The values in the Meas column represent the actual measured value of the
current calibration.
The values in the Drift column represent the difference between the measured
value from the current calibration and the value assigned to the analyte during
the manufacture of the cartridge.
If the system cannot report a value during a calibration, the result is blank on
the report and the D code and explanation prints at the bottom of the report
The following symbols can display with the results on the printed report.
Table 6-1 Symbols or
This symbol... Indicates that...
or the result is above or below the drift limit.

Table of Contents

Related product manuals