6-60 Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
D23 Codes
D23 Reagent Flow Error: 2, 10
D23 Reagent Flow Error: 3, 11
Problem Corrective Action
Wash reagent was
not detected during
calibration.
2 FD1
10 FD2
1. Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
2. Perform a leak test.
Refer to Performing the Leak Test‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed.
For Rapidlab 1245 and 1265 systems, verify that the CO-ox
sample chamber is installed correctly.
3. If the problem persists, replace the wash cartridge.
4. If the problem persists, call for technical assistance.
Problem Corrective Action
200 Cal reagent
was not detected
during calibration.
3 FD1
11 FD2
1. Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
2. Perform a leak test.
Refer to Performing the Leak Test‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
3. If the problem persists, replace the reagent cartridge.
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
4. If the problem persists, call for technical assistance.