C-2 Rapidlab 1200 Operator’s Guide: System Fluids
02087462 Rev. V
Rapidlab 1200 Cartridges
Each of the cartridges has an install-by date on the label.
Cartridge Stability Use life Storage temperature
Reagent
cartridge
30 days Calibration dependent
2
° to 8°C
Room temperature not to
exceed 25°C, up to 7 days
Wash Cartridge
Rapidlab 1245
and 1265
18 days Sample dependent
30 samples per day = 12
days of use
Room temperature at
approximately 25°C
Wash Cartridge
Rapidlab 1240
and 1260
18 days Sample dependent
30 samples per day = 17
days of use
Room temperature at
approximately 25°C
AutomaticQC
cartridge
28 days sample dependent
at least 1 sample analysis
of each level, 3 times per
day
2
° to 8°C