Rapidlab 1200 Operator’s Guide: Maintenance 5-17
02087462 Rev. V
Replacing the Reagent Manifold
You need a reagent manifold for this task.
BIOHAZARD: Wear personal protective equipment. Use universal precautions.
Refer to Appendix A, Protecting Yourself from Biohazards for
recommended
precautions when working with biohazardous materials.
1. Select Status > Diagnostics > Cartridges.
2. Select
Eject Both Cartridges, remove the cartridges, and set them aside.
3. Remove the reagent manifold:
a. Lift the front cover.
Refer to Preparing for Maintenance Procedures‚ page 5-1.
b. Disconnect the waste tubing from the reagent manifold.
c. For Rapidlab 1245 and 1265 systems, disconnect the reagent manifold
tubing from the CO-ox pump tubing at the fitting.
Do not disconnect the reagent manifold tubing from the reagent manifold.
Figure 5-7 Sample Chamber and Reagent Manifold
1 Fitting
2 Reagent manifold tubing (Rapidlab 1245 and 1265 systems)
3 Waste tubing