EasyManua.ls Logo

Siemens Rapidlab 1200 - D70 Optics Error: 9; D70 Optics Error: 11, 12; D71-D77 Codes; D71 no Sample Detected at FD3

Siemens Rapidlab 1200
448 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rapidlab 1200 Operator’s Guide: Troubleshooting 6-71
02087462 Rev. V
D70 Optics Error: 9
D70 Optics Error: 11, 12
D71–D77 Codes
D71 No Sample Detected at FD3
Problem Corrective Action
The system detects
a failure in the
CO-ox optical
components.
1. Select
Status > Shutdown.
2. Wait 10 seconds, and then turn the system on.
3. If the problem persists, call for technical assistance.
Problem Corrective Action
The system detects
a failure in the
CO-ox optical
components.
1. Perform a full calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
2. Select
Diagnostics > CO-ox > Wavelength Cal.
3. Select
Status > Shutdown.
4. Wait 10 seconds, and then turn the system on.
5. If the problem persists, call for technical assistance.
Problem Corrective Action
Fluid detector 3
(FD3) does not
detect the sample
during the
predefined time
limit.
CO-ox results are
not reported. Blood
gas, pH, electrolyte,
and metabolite
results are reported.
Ensure the sample volume is sufficient for the sample type.
Select
Diagnostics > Fluidic Functions > RCx.
Look for obstructions or poor flow in the CO-ox sample
path.
Remove obstructions from the CO-ox sample path:
If the obstruction is in the CO-ox sample tubing, replace the
tubing.
If the obstruction is in the CO-ox pump tubing or waste
tubing, remove the obstruction.
Repeat RCx test to check flow through the sample path.
Deproteinize the sample path.
Refer to Deproteinizing the Sample Path‚ page 5-4.
Perform a full calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
If the problem persists, call for technical assistance.

Table of Contents

Related product manuals