Rapidlab 1200 Operator’s Guide: Maintenance 5-13
02087462 Rev. V
b. Connect the new tubing to the pressure sensor bubbler and to the reagent
manifold.
6. Dispose of the tubing according to your institution’s protocol.
7. Close the reagent cartridge door and then close the front cover.
8. If you have a Rapidlab 1245 or 1265 system and are replacing the CO-ox
tubing, continue with step 2 of Replacing the CO-ox Module Tubing
page 5-13.
9. Close the measurement module door.
10. Select
No.
11. Select
Return > Analyze > Yes.
Replacing the CO-ox Module Tubing
You need a CO-ox tubing kit for this task.
BIOHAZARD: Wear personal protective equipment. Use universal precautions.
Refer to Appendix A, Protecting Yourself from Biohazards for
recommended
precautions when working with biohazardous materials.
1. Select
Status > Diagnostics > Cartridges.
2. Select
Eject W Cartridge and remove the wash cartridge.
3. Lift the door to the measurement module.
Refer to Preparing for Maintenance Procedures‚ page 5-1.