Rapidlab 1200 Operator’s Guide: Maintenance 5-55
02087462 Rev. V
Flushing the Tubing for the Measurement Module
1. Disconnect the tubing from the left side of the pressure sensor bubbler.
2. Connect the syringe tubing to the pressure sensor bubbler.
3. Flush the measurement module tubing with a minimum of 3 mL water.
4. Flush the measurement module tubing with a minimum of 12 mL air.
5. Collect the water in an absorbent swab or equivalent.
6. Reconnect the tubing to the pressure sensor bubbler.
Figure 5-29 Measurement Module Pressure Sensor Bubbler
1 Pressure sensor bubbler
2 Disconnected tubing
3 Water collection area