1-20 Rapidlab 1200 Operator’s Guide: System Overview and Intended Use
02087462 Rev. V
Viewing the Banner Information
The screen banner area displays the following information:
the current system status (Ready, Not Ready, Calibrating, Analyzing)
the time and type of the next calibration
status messages (Cal Pending, Required QC Due, AQC Pending, maintenance due or
overdue)
temperature of patient sample in banner at Results screens, if temperature
demographic is selected in Setup
remote viewing status for systems connected to an LIS (refer to Setting Up Remote
Viewing (Rapidcomm Only)‚ page 8-32)
Figure 1-24 Banner–Status Information
Viewing Status Messages
The banner displays status messages as reminders for pending tasks. The system always
displays a message for a pending Required QC regardless of other pending tasks.
The system displays a message when maintenance tasks are due and how many are due.
When you perform a maintenance task, the number of tasks on the banner decrements but
the message remains as long as tasks are pending.
Status Symbols
When a cartridge, the CO-ox sample chamber, or the waste bottle approaches its
expiration date or number of tests available, the system displays the appropriate symbol in
the banner:
1 System status area
2 Cartridge status
Supply Symbol % Volume Number of hours
Wash cartridge 10% Less than 24
Reagent cartridge 10% Less than 24
AutomaticQC cartridge 10 or fewer tests for
any level of QC
material
Less than 24