Rapidlab 1200 Operator’s Guide: System Configuration 8-23
02087462 Rev. V
7. If you select a Calibration Report, select either Status or Full.
Refer to Generating Calibration Reports‚ page 3-3.
8. If you are selecting a Sample or QC report, select
Ranges to include ranges
on the report.
The default for including ranges is off.
9. If you want to print the report to the other printer, select it and make the
selections you want for the report on that printer.
10. Select another report, as needed.
When you select another report, the system displays the options for that
report. You must repeat the above procedure to set up printing for each report
that you want to print.
11. When you are done selecting reports, select
Save.
Setting up Barcode Options
Use this screen to turn the barcode scanner on or off, and to specify the symbology
and format for patient ID, accession number, and password barcodes.
This screen can also be used to select the barcode only option for Patient ID data
entry. Refer to
Selecting the Barcode Only Option for Patient ID Entry‚
page 8-25.
• The patient ID can be up to 20 alphanumeric characters in length. The
accession number and password barcodes can be up to 13 alphanumeric
characters in length.
• Within each barcode, you can turn any characters off that you do not want the
system to include as part of the patient ID or the accession number.
• The system automatically turns off any non alphanumeric characters, such as
hyphens, that are displayed in the barcode.
• You can specify 1 or 2 fixed lengths for Interleaved 2 of 5 barcodes.
NOTE: If the barcode scanner was not connected during installation of the
system, connect the barcode scanner to the system. Refer to Connecting the
Barcode Scanner‚ page 8-26, before completing this procedure.
1. At the Setup screen, select
Printer and Devices > Bar Code Setup.
NOTE: If you have selected the Required QC analysis option, you must leave
the barcode scanner on.
2. You can select
barcode to turn the barcode scanner on or off.