EasyManua.ls Logo

Siemens Rapidlab 1200 - Appendix E: System Specifications; Viewing Parameter Measurements

Siemens Rapidlab 1200
448 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
02087462 Rev.V
Appendix E: System Specifications
The following table lists the specifications for the Rapidlab 1200 systems.
Viewing Parameter Measurements
The following table lists the units, reporting ranges, and display resolutions for the
pH and blood gas parameters measured by the Rapidlab
1200 system.
Table E-1 Rapidlab 1200 System Specifications
Property Specification
Ambient operating temperature 15–32°C
Ambient temperature change less than 4°C per hour
Ambient operating relative humidity 5–90%, non-condensing
Ambient shipping temperature -25–45°C
Ambient storage temperature 4–45°C
Power rating 150VA
Voltage requirements 100V/240V (85V to 264V autoranging)
50/60Hz
Ambient operating barometric
pressure
523–800 mmHg (69.7–106.7 kPa)
System dimensions, base model
height 61.0 cm (24 inches)
*
width 53.5 cm (21 inches)
depth 58.5 cm (23 inches)
weight 29.5 kg (65 lbs)
* Height with screen fully upright
Table E-2 pH and Blood Gas Parameters
Parameter Units Reporting Range Resolution
pH pH 6.000–8.000 0.001
H
+
nmol/L 10.0–1000.0 0.1
pCO
2
mmHg
kPa
5.0–250.0
0.67–33.33
0.1
0.01
pO
2
mmHg
kPa
0–800
0.0–106.67
0.1
0.01
pAtm mmHg
kPa
523–800
69.7–106.6
1
0.1

Table of Contents