2-20 Rapidlab 1200 Operator’s Guide: Operating the System
02087462 Rev. V
Using Custom Panels
When you define custom panels in Setup, the Analysis screen displays a set of custom
panels in the lower left of the screen. For procedures to define custom panels, refer to
Defining Custom Panels‚ page 8-11.
Figure 2-6 Custom Panel Buttons
Custom panels defined in Setup display at the Analysis screen.
If at least 1 panel is defined in each of the 2 custom panel sets, buttons labeled 1 and 2
display at the Analysis screen. Use these buttons to select Set 1 or Set 2. Each set
contains 1-3 panels.
You cannot view the custom panels in Set 1 and Set 2 at the same time. Selecting one
set at the Analysis screen deselects the other set.
The Analysis screen only displays a custom button if you select parameters for that
panel. For example, if you only select parameters for 2 of the 3 panels in Set 1, only
those 2 panels display when you select Set 1.
If you disable a parameter in Setup after selecting a custom panel that contains that
parameter, the system automatically removes that parameter from that panel.
Only patient samples can be used with custom panels.
To use QC, AQC, or High G/L samples, deselect the custom panel.
Only parameters for the selected custom panel display at the Analysis screen. To
display other parameters, you must deselect the custom panel.
NOTE: If a parameter on a selected custom panel is outside the defined QC range, that
parameter remains selectable so you can restore it. Refer to Restoring Parameters
page 4-12.
1 Custom panel buttons
2 Set buttons