Rapidlab 1200 Operator’s Guide: System Configuration 8-19
02087462 Rev. V
Setting up Values
Use this screen to enter the default values for the O
2
Binding Factor (OBF) and
ctO
2
(a- ):
1. At the Setup screen, select
Parameters > Values.
The screen shows the current values.
2. Select the demographic you want to change.
3. Enter a value and then select
Save.
4. Select
Save.
Setting up System Options
When you select System Options at the Setup Window, the following options are
available:
Country Options
Date and Time
Sound
•Other
Setting up Country Options
The Country Options screen allows you to perform the following tasks:
Select a language for the system
screens and videos
Select a date format for your laboratory
O
2
ctO
2
(a- ), t
* If ctHb is not available as an entered value and tHb is not available, the system uses 15 g/dL as a
default value.
The system uses a default value for these parameters.
A calculated value determined from the total hemoglobin value.
Parameter Required Parameters and Sample Demographics