5-46 Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
4. Grasp the lower tubing cuff and gently pull it away from the platen.
5. While holding the lower side of the tubing, turn the roller cage clockwise and gently
pull the tubing away from the platen and the roller cage.
Figure 5-23 CO-ox Module
6. Gently pull the roller cage straight off its shaft.
7. Dispose of the roller cage according to your institution’s protocol.
Cleaning the Roller Cage Shaft
Clean the roller cage shaft using lint-free tissues moistened with water and then
thoroughly dry the shaft.
Installing the New Roller Cage
1. Place the new roller cage on the shaft.
2. Press the roller cage down and turn the cage clockwise until the cage snaps into place.
3. Place the lower tubing cuff under the lower side of the platen.
NOTE: Do not stretch the tubing.
1 Platen
2 Roller cage
3 Lower tubing cuff