Rapidlab 1200 Operator’s Guide: Maintenance 5-25
02087462 Rev. V
Cleaning the CO-ox Roller Cage
Use this procedure to clean the roller cages of the CO-ox pump.
You need the following materials for this task:
• Lint-free tissue and swabs
• 10% bleach solution
•Water
1. Select
Status > Maintenance.
2. Lift the front cover.
Refer to Preparing for Maintenance Procedures‚ page 5-1.
3. Disconnect the reagent manifold tubing from the CO-ox pump tubing at the
fitting.
Figure 5-12 CO-ox Pump Tubing to Reagent Manifold
4. Grasp the lower tubing cuff and gently pull it away from the platen.
BIOHAZARD: Wear personal protective equipment. Use universal precautions.
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
1 Fitting
2 Reagent manifold tubing