4-8 Rapidlab 1200 Operator’s Guide: Quality Control
02087462 Rev. V
Using the Recall QC Data Search Screen
The Recall QC Data Search screen allows you to search for a specific QC lot, QC sample
results performed between a defined date range, or a combination of criteria. The QC Data
Search screen acts as a filter to display only those results that meet the criteria you specify
on the Search screen.
1. At the QC List screen, select Search.
2. At the search screen, use the following search options to set your criteria:
NOTE: Use the number pad to enter your criteria where necessary.
3. When you are done selecting the criteria, select
Continue.
Select this option... To do this...
Lot enter the lot number of the control you
want to view.
If you do not enter a lot number, the
system displays all lots that meet the other
criteria you define.
Start Date enter the earliest date for search on a range
dates.
The start date must be earlier than the End
Date.
End Date enter the latest date for a search on a range
of dates.
If you do not enter an end date, the system
uses the current day date.
Level 1, 2, 3 display QC results according to level.
You may use any combination of levels or
select all three.
Unknown to display samples when levels are not
identified.
High G/L to select only High G/L results.
You can use this in combination with the
other options.