Rapidlab 1200 Operator’s Guide: Troubleshooting 6-45
02087462 Rev. V
Using the CO-ox Screen
Only Rapidlab 1245 and 1265 systems use the functions on the CO-ox screen.
The Diagnostics–CO-ox screen displays buttons for performing the following
actions:
testing the lamp calibration
testing wavelength calibration
testing the CO-ox Sample Chamber
turns the CO-ox lamp off or on
Performing the Lamp Calibration Test
This test is only performed on Rapidlab 1245 and 1265 Systems.
Perform this test whenever you change the CO-ox lamp or whenever you remove
and reinstall the lamp. You can also use this test to perform a lamp calibration.
The system displays the pass or fail results on a printed report.
1. Select Status > Diagnostics.
2. Enter your password, if necessary.
3. Select
CO-ox > Lamp Cal.
4. Wait while the system performs the test.
The system prints test results as Passed or Failed.
5. Select
Return.
6. To return to the Status screen, select
Return.
The system performs a full calibration.