Rapidlab 1200 Operator’s Guide: Troubleshooting 6-3
02087462 Rev. V
Troubleshooting the Purple Parameter Error
Troubleshooting the Ampule QC
Use the following recommendations to help resolve issues with failed QC
performed from an ampule:
• Perform only one QC analysis for each ampule.
Performance claims are based on one QC analysis per ampule.
• Always use the Quick adapter when analyzing QC samples.
• Analyze the QC material immediately after opening the ampule.
Prolonged contact with the atmosphere affects pH, pCO
2
, and pO
2
results.
• Ensure that the correct target ranges are defined in Setup.
• Store QC material at a constant temperature.
• Do not handle ampules excessively.
Variation in results may occur due to changes in temperature.
• Do not use expired QC materials.
• Do not analyze QC materials that contain bubbles.
• Use the correct level of QC material that is scheduled for analysis.
Error
Probable
Cause Corrective Action
The parameter
is purple with a
line through it
and QC! is
displayed in
the lower left
corner.
Required QC
analysis was
not performed
when
scheduled.
• Analyze all the currently scheduled controls.
Refer to Performing Required QC Sample
Analysis‚ page 4-1.
• If the scheduled controls do not contain the
necessary levels, analyze QC for the
appropriate levels.
Refer to Performing Unscheduled QC Sample
Analysis‚ page 4-4.
• If the results are within range, turn the
parameter on.
Refer to Restoring Parameters‚ page 4-12.